Skip to content

Bug: Conductivity errors for Blynken & Wynken #141

@oceanicsdotio

Description

@oceanicsdotio

Seeing the same intermittent drops in conductivity that happened during the harbor test. This is probably wiper parking behavior as noted elsewhere. Need to cross reference and see if it is the same sonde/Ai1 and possibly adjust the wait time after wiping to ensure that it is complete and does not interfere with measurement. Might be worth doing a bench test to replicate after the buoy is recovered, to make sure that is the root cause. The buoy that is staged to be deployed at Hurricane had improperly seated connector for the wiper, but I checked to make sure that didn't happen on this deployment.

Metadata

Metadata

Labels

bugSomething isn't workingdataRelated to analyzing or using datasetselectronicsDealing with electronicssensorsOther sensors

Type

Projects

No projects

Relationships

None yet

Development

No branches or pull requests

Issue actions